Gendex GSX-700 within Eaglesoft

To get the Gendex GSX-700 Inter-oral Xray sensor to work with eaglesoft, complete the following 3 step process:

Step 1 :
Check the version of GxPicture the office is currently using.
Download the GXPicture version that the office is currently using from http://www.gendex.com/software-drivers
Install GXPicture
choose other with filters
Select the xray sensor type on the next page and hit enter

Step 2 :
Run this registry hack so the sensor/camera will show up in Eaglesoft:

32 Bit Machine: Regedit : Hkey_Local_Machine / Software / Eaglesoft / Chairside / Digital Xray
64 Bit Machine: Regedit : Hkey_Local_Machine / Software/ Wow6432 / Eaglesoft / Chairside / Digital Xray

New Dword 32 bit value, call it GXS Installed Modify > Value of 1

 

Step 3 :
Install calibration/correction files

Copy the Gendex Calibration files from another Operator y room by going to following locaton.

C:\Program Files or Progam Files X86 \ Gendex \ (XXXX)_Calibration Files

Copy the entire (XXXX)_Calibration Files folder to new workstation same location.

Step4
Update Eagelsoft Preference Setting Xray tab

Choose default device as sensor and choose Gendex Imaging as sensor.

Step 5

Make sure Gendex icon in task manager shows green when sensor is plugged in.

Take a test xray or have the practice take a test xray to confirm imaging is working.